Navigation Horizontal | Navigation Vertical / Unternavigation | Direkt zum Inhalt
 
Header

Inhalt:

Liquid Filters Overview

Catalogue-
No.
Use Consisting of Substance Wavelength nm
667.003-UV Set for testing photometers  according to Ph. Eur.
667.100-UV
667.200-UV
667.305-UV
667-UV5
Potassium chloride and reference filter
Toluene in hexane and reference filter
Potassium dichromate in HClO4 and reference filter Holmium oxide in perchloric acid
198, 200 (certified)
266, 269
235, 257, 313, 350, 430
241,15; 287,15; 361,5; 536,3

All sets are also available individually!

667.100-UV Set for testing for stray light according to Ph. Eur. 667-UV1
667-UV12
Potassium chloride
Purified water (reference filter)
200 (Cut-Off) 198, 200 (certified)
667.200-UV Set for testing the spectral resolution according to Ph. Eur. 667-UV6
667-UV9
Toluene in hexane
Hexane (reference filter)
266, 269
667.305-UV Set for testing the photometric accuracy according to Ph. Eur. 667-UV60
667-UV600
667-UV14
Potassium dichromate in HClO4
Potassium dichromate in HClO4
(430 nm)
Perchloric acid (reference filter)
235, 257, 313, 350
430
667-UV5 Testing the wavelength accuracy according to Ph. Eur. 667-UV5 Holmium oxide in perchloric acid 241,15; 287,15; 361,5; 536,3

667.101-UV Set for testing for stray light 667-UV10
667-UV12
Sodium iodide
Purified water (reference filter)
259 (Cut-Off)
220 (certified)
667.102-UV Set for testing for stray light 667-UV11
667-UV12
Sodium nitrite
Purified water (reference filter)
385 (Cut-Off)
340, 370 (certified)
667.307-UV Filter set for testing the absorption and linearity 667-UV20
667-UV40
667-UV60
667-UV80
667-UV100
667-UV14

Potassium dichromate in perchloric acid
Perchloric acid (reference filter)
235, 257, 313, 350
667.400-UV Set for testing the wavelength accuracy 667-UV5
667-UV14
Holmium oxide in perchloric acid
Purified water (reference filter)
241.15, 287.15,
361.5, 536.3