Navigation Horizontal | Navigation Vertical / Unternavigation | Direkt zum Inhalt
hellma analytics

Inhalt:

Didymium Glass Filter

Description Didymium glass filter, solid filter, certified
Use Assessment of the photometric accuracy in the UV region (270 nm to 340 nm)
Material Glass BG 36 of SCHOTT Glaswerke

666-F7 Didymium glass filter

Kalibrierstandards

The F7 marked didymium glass filter is actually a standard to check for wavelength accuracy in the UV region, but it can also be used for checking the photometric accuracy.

Checking the photometric accuracy in the UV range is possible at the following wavelengths: 270 nm, 280 nm, 297 nm, 320 nm and 340 nm. As a standard, the thickness of the filter is manufactured in a way that it will have an absorbance of approx. 0.5 A at a wavelength of 340 nm. Absorbance values will increase with decreasing wavelengths up to ca. 1.5 A at 270 nm.

E_Didymiumglas-Filter Typical spectrum of a Didymium glass filter between 270 nm and 370 nm

Catalogue-
No.
DescriptionMaterialThick-
ness mm
Nominal AbsorbanceAbsorbance
value
at nm
666-F7Didymium Glass Filter,
Certified
BG 36ca. 1,2depending on
wavelength
270, 280, 297, 320, 340

Weitere Informationen:

 

Contact

Hellma Analytics
Klosterrunsstraße 5
79379 Müllheim
Germany
Phone +49 7631/182-1010
Fax. +49 7631/182-1011

sales.analytics@hellma.com