Potassium dichromate in Perchloric Acid (430 nm)

Description Potassium dichromate filter, liquid filter, certified
Use Assessment of photometric accuracy (absorbance) in the Vis range (measuring wavelength 430 nm) with a spectral bandwidth of 2 nm or less
Material 0.06% potassium dichromate dissolved in perchloric acid and sealed in Hellma precision cells made from quartz SUPRASIL®

667-UV600 0.06% Potassium dichromate dissolved in perchloric acid

HA_CAL_667-UV600-430nm_Totale_12

Fifth revision of European Pharmacopeia stipulates also checking the photometrical accuracy of the spectrophotometer at a wavelength in the visible region (430 nm) besides the checking in the UV region. The testing has to be carried out by means of a solution of 0.06% potassium dichromate dissolved in perchloric acid. The absorbance value of the filter is determined at a wavelength of 430 nm and is specified in the certificate. It should be noted that the measurement is carried out with the filter 667-UV14 (0.01 N perchloric acid) in the reference beam path. Measured values of the reference filter are separately displayed.

E_ Kaliumdichromat-Flüssigfilter 430 nm Typical spectrum of a 0.06 % potassium dichromate solution

Wavelength
(nm)
Specific Absorption
A 1 %1 cm
Maximum Tolerance
A 1 %1 cm
430 15,9 15,7 to 16,1