Certified Reference Materials to Check the Photometric Accuracy

To check the photometric accuracy Hellma Analytics offers different kinds of standards: neutral density glass filters are very well suited for testing in the visible range, checking the photometric accuracy with potassium dichromate dissolved in perchloric acid is the approved method for the UV region. Typically used reference materials have broad peaks and valleys. The following table shows the standards offered by Hellma and their compliance references:

Checking for Calibration Standard Pharmacopeia ASTM
EP DAB USP
Photometric
Accuracy
Potassium Dichromate Liquid Filter X X X X
Neutral Density Solid Filter X X
Didymium Glass Solid Filter